Flexray Decoding (Option) - Rigol MSO8000A Series User Manual

Digital oscilloscope
Hide thumbs Also See for MSO8000A Series:
Table of Contents

Advertisement

Protocol Decoding
- CRC (Cyclic Redundancy Check): expressed in Hex, displayed in yellow-green.
When errors occur, it is displayed as a red patch.
- ACK (Acknowledgement): when effective, displayed as a light yellow-green
patch; when errors (ACK is detected to be 1) occur, displayed as a red patch.
- R (remote frame): displayed as a tangerine patch.
- Stuff (Bit filling error): displayed as a red patch.
15.7

FlexRay Decoding (Option)

FlexRay is a type of differential serial bus configured with three consecutive segments
(i.g. packet header, payload, and packet trailer). The oscilloscope samples the FlexRay
signal at the specified sample position and judges each data point as logic "1" or
logic "0" according to the preset threshold level. The FlexRay decoding is required to
specify the signal type and baud rate.
In the decode setting menu, press Decode1 > Bus Type, then rotate the
multifunction knob
also press Bus Type continuously or enable the touch screen to select it.
1. Enable or Disable the Bus
Press Bus Status continuously to enable or disable the decoding function.
2. Quickly Apply FlexRay Trigger Settings to FlexRay Decoding
Press Copy Trig to copy the current FlexRay trigger settings and apply them to
FlexRay decoding function (set the corresponding FlexRay decoding parameters
automatically). For analog channels, the threshold level settings will also be copied.
3. Set the Source and the Threshold
- Press Source, rotate the multifunction knob
channel, and then press down the knob it. You can also press Source
266
Frame ID
DLC
to select "FlexRay". Press down the knob to select it. You can
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Data
CRC
to select the desired source
ACK

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mso8074aMso8154aMso8204a

Table of Contents