- Press MOSI, then rotate the multifunction knob
5. Set the Data Parameters
Press Settings to enter the data setting menu.
- Set the endian
Press Endian to select "LSB" or "MSB". By default, it is MSB
LSB: indicates Least Significant Bit transmission sequence, i.g. the lowest bit of
the data is transmitted first.
MSB: indicates Most Significant Bit transmission sequence, i.g. the highest bit of
the data is transmitted first.
- Set the polarity
Press Polarity continuously to select Positive
for data decoding.
- Set the width
Press Width, then rotate the multifunction knob
keypad to set the bits of each frame of data. The setting range is from 4 to 32.
By default, it is 8.
6. Set Display-related Parameters
Press Display to enter the display setting menu.
- Set the display format
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
MISO, then rotate the multifunction knob
set the threshold of the MISO data channel.
channel, and then press down the knob to select it. You can also press MOSI
continuously or enable the touch screen to select it. The available channels
include OFF, analog channels (CH1-CH4), and digital channels (D0-D15).
When you select the analog channel (CH1-CH4), press Threshold below
MOSI, then rotate the multifunction knob
set the threshold of the MOSI data channel.
NOTE
The source channels of MISO and MOSI cannot be set to "OFF" at the same time.
Protocol Decoding
or use the numeric keypad to
to select the MOSI data
or use the numeric keypad to
or Negative
as the polarity
or use the pop-up numeric
251