Trigger Condition:
Press When, then rotate the multifunction knob
and then press down the knob to select it.
You can also press When continuously or enable the touch screen to select it.
Sync: triggers on the specified sync type. After this trigger condition is selected,
press Sync to select the desired sync type: Data Sync, Cmd/Status Sync, or All
Sync.
Data: triggers on the specified data word. After this trigger condition is selected,
press Comp to select the comparison conditions. The available choices include
=, ≠, >, <, ><, and <>.
-
-
-
-
-
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
=: triggers when the channel's data word equals the set data word. Press
Min to set the lower limit of the data word. For setting methods, refer to
I2C Trigger (Option)
descriptions in "
≠: triggers when the channel's data word does not equal the set data word.
Press Min to set the lower limit of the data word. For setting methods, refer
I2C Trigger (Option)
to descriptions in "
<: triggers when the channel's data word is smaller than the set data word.
Press Max to set the upper limit of the data word. For setting methods,
refer to descriptions in "
>: triggers when the channel's data word is greater than the set data word.
Press Min to set the lower limit of the data word. For setting methods, refer
I2C Trigger (Option)
to descriptions in "
< >: triggers when the channel's data word is smaller than the upper limit
of the data word and greater than the lower limit of the data word. Press
Max and Min to set the upper limit and lower limit of the data word
respectively. For setting methods, refer to descriptions in "
(Option)
".
to select the trigger condition,
".
".
I2C Trigger (Option)
".
".
To Trigger the Oscilloscope
I2C Trigger
131