Quick Start
5. Observe the waveform on the display. In normal condition, the square waveform as
shown in
6. Use the same method to test the other channels. If the square waveforms actually
shown do not match that in the figure above, please perform "
Compensation
WARNING
To avoid electric shock when using the probe, please make sure that the insulated wire of
the probe is in good condition. Do not touch the metallic part of the probe when the
probe is connected to high voltage source.
4.3.7
Probe Compensation
When the probes are used for the first time, you should compensate the probes to
make them match the input channels of the oscilloscope. Non-compensated or
poorly compensated probes may cause measurement inaccuracy or errors. The probe
compensation procedures are as follows:
1. Perform Step 1, 2, 3 and 4 specified in "
2. Check the displayed waveforms and compare them with the waveforms shown in
Figure 4.12
20
Figure 4.11
should be displayed.
Figure 4.11 Square Waveform Signal
" introduced in the next section.
.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Probe
Function Inspection
".