Start Sample Measurement With The Wizard; Use Existing Data; Report Test Results; How To Use The Print Center - Beckman Coulter MET ONE 3411 User Manual

Table of Contents

Advertisement

b. Highlight a location. Use the UP or DOWN arrows to change the
order of locations as needed. The order of the list will be the
order of the test.
6. Go to the next screen. A confirmation of the selections will show.

Start sample measurement with the Wizard

1. Push BEGIN SAMPLING.
2. Obey the prompt and go to the first sample location. Push OK.
3. Put the isokinetic probe in position for the test. Push RUN.
4. Continue to obey the prompts and move to each location for the test.
When all tests are complete, the counter will return to the wizard to
configure the report.

Use existing data

The Test and Report Wizard can include existing date for the selected
standard and location.
1. Select the Use Existing Data checkbox.
2. Enter the date range for the existing data.
3. Push BEGIN SAMPLING.

Report test results

• Push
to send the selected data to a USB drive. Refer to
data
on page 36.
• Push
to print the selected data. Refer to
Center
on page 34.

How to use the Print Center

About the Print Center

N O T I C E
To avoid damage to the printer, do not operate the printer without paper. If the
particle counter must be used without paper, be sure to set the Sample Print
Mode field to "None".
34 English
Export
How to use the Print
The particle counter has a built-in printer. The Print Center screen
is accessible from the:
• Counter Navigation screen
• Historical screen
• System Diagnostics screen
• Test/Report Wizard screen
• Area/Location Setup screen
On the Print Center screen, the user can:

• Set automatic print functions

• Print buffer records or count averages
Note: Filtered data is printed from the Historical screen.
Set automatic print functions
Note: If the sample period is very brief and the hold time is zero, the sending of
some sample data to the printer may be skipped. However, the data is always
obtained and saved. To avoid a printout failure, set the total time of the Sample
period and the Hold period combined to more than 12 seconds.
1. On the Counter Navigation screen, push PRINTER.
2. On the Print Center screen, select the Sample Print Mode field.
Select an option for automatic printing.
Option Description
None
No data will print automatically
Alarms Prints results at the end of the sample when a count alarm is
exceeded
Cycles Prints the results of multiples of the programmed count cycle. The
range is 1–99.
All
Prints results after each count cycle is finished
3. Set the Cycle Print Order. Options: 10–1 (oldest printed first) or
1–10 (newest printed first).