P R E L I M I N A R Y
D
FR091521, R
IRECTION
8-5-12
CFE Release Arm Replacement Procedure
8-5-12-1
Tools
Appropriate Phillips screwdriver.
FRU Part #
Refer to
8-5-12-2
Time Required
15 minutes
8-5-12-3
Preparation
Shut the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
8-5-12-4
CFE Board Release Arm Removal Procedure
The lever-shaped CFE Release Arm is located on the PSB Board - see
1) Remove the CFE Board:
•
2) Uninstall the CFE Release Arm (Lever) by unscrewing the two securing screws as indicated
in
8-5-12-5
CFE Release Arm Installation Procedure
1) Install a new CFE Release Lever in position on the PSB and secure with the two securing screws
previously removed.
2) Refit the CFE Board:
•
8-116
Section 8-5 - Electronic Cage Components - Replacement Procedures
1
EVISION
Table 9-16
on page 9-15
Figure 8-142 Location of CFE Release Arm
CFE Board Removal Procedure on page 8 - 112
Figure 8-142,
above.
CFE Board Installation Procedure on page 8 - 115
.
V
S60N/V
S70N
IVID
IVID
Power Shut Down on page 4 -
Figure
B
S
M
ASIC
ERVICE
ANUAL
7.
8-142.