Bios Battery Replacement; Tools; Time Required; Preparations - GE VIVID S60N Basic Service Manual

Hide thumbs Also See for VIVID S60N:
Table of Contents

Advertisement

P R E L I M I N A R Y
D
FR091521, R
IRECTION
8-5-5

BIOS Battery Replacement

NOTE:
Since the BIOS battery is not available as a spare part (due to shelf life), it should be obtained locally.
The battery model is CR2032.
8-5-5-1

Tools

Use the appropriate Phillips screwdriver.
8-5-5-2

Time Required

1 min
8-5-5-3

Preparations

Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
BEFORE REMOVING THE BEP, REMEMBER TO PERFORM A FULL INCREMENTAL BACK-UP OF
CAUTION
THE ARCHIVE AND USER PRESETS.
For details, refer to the
page 8 -
8-5-5-4

BIOS Battery Removal Procedure

1) Remove the BEP:
2) Place the BEP on a clean, secure surface.
Note:
The location of the BIOS battery in the BEP is shown in
8-98
Section 8-5 - Electronic Cage Components - Replacement Procedures
1
EVISION
Operating System and/or Application Software Loading Procedures on
173.
BEP Removal Procedure on page 8 - 94
Figure 8-122 Location of BEP BIOS Battery
V
S60N/V
S70N
IVID
IVID
Power Shut Down on page 4 -
Figure
8-122.
B
S
M
ASIC
ERVICE
ANUAL
7.