CPU Partition Replacement Procedure
NOTE:
The CPU partition is located on the rear side of the BEP- see
8-5-8-1
Tools
Appropriate Phillips and flat screwdrivers.
FRU Part #
Refer to
8-5-8-2
Time Required
20 min
8-5-8-3
Preparations
Shut down the Vivid S60/Vivid S70 ultrasound unit, as described in
8-5-8-4
CPU Partition Removal Procedure
1.) Remove the BEP:
•
2.) Place the BEP on a clean work surface with the CPU Partition facing upwards.
3.) Remove the two securing screws (shown in
4.) Gently slide the CPU Partition upwards and remove it from the BEP.
8-104
Section 8-5 - Electronic Cage Components - Replacement Procedures
1
EVISION
Figure 8-129 CPU Partition on BEP
Table 9-14
on page 9-13
BEP Removal Procedure on page 8 - 94
.