Cooling System; General Information - GE VIVID S60N Basic Service Manual

Hide thumbs Also See for VIVID S60N:
Table of Contents

Advertisement

P R E L I M I N A R Y
D
FR091521, R
IRECTION
Section 5-12

Cooling System

5-12-1

General Information

The Vivid™ S60/Vivid™ S70 system has 3 fans for system cooling:
All fans are controlled and have variable speed.
The cooling requirement for the Vivid™ S60N/Vivid™ S70N ultrasound scanner with monitor and on
board peripherals, is up to 2000 BTU/h. This figure does not include cooling needed for lights, people,
or other equipment in the room.
NOTE:
Each person in the room places an additional 300 BTU/h demand on the cooling system.
5-28
1
EVISION
Back End Processor Fan
Front End Fan
AC Distribution Box Fan
Section 5-12 - Cooling System
V
™ S60N/V
IVID
- located at the rear of the electronic cabinet.
- located on the metal box of the TRX32 Boards (on the cFE).
- located on the side of the AC Distribution Box.
™ S70N P
IVID
ROPRIETARY
S
M
ERVICE
ANUAL