Front End Power Supply Replacement Procedure
NOTE:
The Front End Power Supply (FEPS) is also referred to as the HVPS (High Voltage Power Supply).
CAUTION
8-5-10-1
Tools
Appropriate Phillips screwdriver.
FRU Part #
Table 9-16
Refer to
8-5-10-2
Time Required
15 minutes
8-5-10-3
Preparation
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
8-5-10-4
Front End Power Supply Removal Procedure
1) Remove the following covers: left side, right side, DVD, front.
•
Air Inlet (Left Side) Cover Removal Procedure on page 8 - 7
•
Right Side Cover Removal Procedure on page 8 - 10
•
DVD Cover Removal Procedure on page 8 - 11
•
Front Cover Removal Procedure on page 8 - 16
2) Open and remove the Front End door assembly and place it on a stable surface:
•
Front End Door Assembly Opening and Removal Procedure on page 8 - 83
3) Loosen the 2 retaining screws that secure the FEPS to the Front End door assembly and gently lift
the plate away from the guide pins located on either side of the FEPS as shown in
Guide Pin
1
When performing these procedures, take precautions to avoid damage of
electrostatic-sensitive components. Always have the ESD wrist strap
connected either to the DIB chassis or to the GND plug at the rear of the
scanner, and to your hand.
If a battery is present, first remove the battery as it contains stored energy.
Refer to
Battery Removal Procedure on page 8 -
.
on page 9-15
Figure 8-133 Removing the FEPS Metal Cover