Mechanical Functions Checks; Monitor Articulated Arm Movement Check; Operator Panel Movement Check - GE VIVID S60N Basic Service Manual

Hide thumbs Also See for VIVID S60N:
Table of Contents

Advertisement

P R E L I M I N A R Y
D
FR091521, R
IRECTION
4-3-14

Mechanical Functions Checks

4-3-14-1

Monitor Articulated Arm Movement Check

Table 4-5
Step
1
2
3
4
4-3-14-2

Operator Panel Movement Check

Table 4-6
Step
1
2
3
4-40
1
EVISION
Monitor Articulated Arm Movement Check
Task
Lift the up/down release handle upwards. (It is located
on the right side below the operator panel).
While lifting the release handle, raise the operator panel
upwards, then downwards and make sure it is able to
reach its maximum lowest and maximum highest
positions.
Release the up/down handle.
Check all positioning capabilities of the Articulated Arm,
as illustrated in
Figure 5-6 on page
Operator Panel Movement Check
Task
Lift the left/right release handle located on the left side of
the Operator Panel.
While lifting the release handle, swivel the operator
panel left and right and make sure it is able to reach its
maximum left and maximum right positions.
Release the left/right handle.
Section 4-3 - Functional Checks
V
S60N/V
IVID
The up and down movement locking mechanism is
released, allowing the arm to be raised or lowered.
Ensure that you do not apply too much force to move
the operator panel and that the movement is smooth.
Ensure that you do not apply too much force to move
the release handle and that the movement is smooth.
During the movement up and down make sure the
image displayed on the monitor does not present any
disturbance.
Make sure the arm is locked and no movement is
observed when moderate force is applied to the
operator panel (upwards and downwards).
Make sure the arm can be moved freely into the
5-9.
illustrated positions.
The left and right movement locking mechanism is
released, allowing the Operator Panel to be swiveled
30 degrees to the right and left from the center position.
Ensure that you do not apply too much force to move
the operator panel and that the movement is smooth.
Ensure that you do not apply too much force to move
the release handle and that the movement is smooth.
During the movement up and down make sure image
displayed on monitor does not present any disturbance.
Make sure Operator Panel is locked and no movement
is observed when moderate force is applied to the
operator panel (left and right).
S70N B
S
IVID
ASIC
ERVICE
Expected Result(s)
Expected Result(s)
M
ANUAL

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivid s70n