Probe/Connectors Check - GE VIVID S60N Basic Service Manual

Hide thumbs Also See for VIVID S60N:
Table of Contents

Advertisement

P R E L I M I N A R Y
D
FR091521, R
IRECTION
4-3-8

Probe/Connectors Check

NOTE:
Probes can be connected at any time, whether the unit is ON or OFF
CAUTION
Table 4-2 Probe and Connectors Checks
Step
Press Probe on the Operator Panel.
1
If not already selected, use the trackball to select the
2
desired probe.
Trackball to the desired application.
Press Select to launch the application.
3
To change application without changing the current
probe, press Appl. on the Operator Panel.
Verify no missing channels.
4
Verify there's no EMI/RFI or artifacts specific to the
5
probe.
Check the probe in each active connector slot.
6
Do a leakage test on the probe.
7
Repeat this procedure for all available probes.
8
Related information:
Electrical Safety Tests on page 10 - 15
4-34
1
EVISION
Take the following precautions with the probe cables:
Keep away from the wheels.
Do not bend.
Do not cross cables between probes.
Task
Section 4-3 - Functional Checks
V
S60N/V
S70N B
IVID
IVID
Expected Result(s)
A list of the connected probes will pop
up on the screen.
An application menu for the desired
probe is listed on the screen.
The selected application starts.
All channels are functioning.
No EMI/RFI or artifacts.
It will display pictorial data each time.
It passes the test.
S
M
ASIC
ERVICE
ANUAL