Power Stability Requirements

Voltage drop-out
Max 10 msec
Power Transients
The Vivid™ S60N/Vivid™ S70N ultrasound scanner is fully compliant with the following standard:
EN/IEC 60501-1-2 sections 36.202.1j, 7a(2).
2-2-6

EMI Limitations

Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies,
magnetic fields, and transients in the air or wiring. They also generate EMI. The Vivid S60N/Vivid S70N
ultrasound unit complies with limits as stated on the EMC label. However, there is no guarantee that
interference will not occur in a particular setup.
Note:
Possible EMI sources should be identified before the unit is installed, and should not be on the
same line as the ultrasound system. A dedicated line should be used for the ultrasound system.
Electrical and electronic equipment may produce EMI unintentionally as the result of a defect. Sources
of EMI include the following:
Medical lasers.
Scanners.
Cauterizing guns.
Computers.
Monitors.
Fans.
Gel warmers.
Microwave ovens.
Portable phones.
Broadcast stations and mobile broadcasting machines.