Mechanical Platform Components - Replacement Procedures

8-6-1

Mechanical Assembly Base Console Replacement

NOTE:
In the event that the Mechanical Base Console requires replacement, contact the OnLine Center for
instructions.
8-6-2

Column Main Support Assembly Replacement

NOTE:
In the event that the Column Main Support Assembly requires replacement, contact the OnLine Center
for instructions.
8-6-3
Mechanical Assembly Interface Base for the Operator Panel Replacement
NOTE:
In the event that the Mechanical Assembly Interface Base for the Operator Panel requires replacement,
contact the OnLine Center for instructions.
8-6-4

Touch Screen Chassis Replacement Procedure

NOTE:
The Touch Screen Chassis complete with brackets are shown in
8-6-4-1

Tools

Phillips screwdriver.
FRU Part #
Table 9-5
Refer to
8-6-4-2

Preparation

Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
8-6-4-3

Time Required

45 mins
1
EVISION
Figure 8-178 Touch Screen Chassis
.
on page 9-3
Chapter 8 - Replacement Procedures
V
S60N/V
S70N
B
IVID
IVID
Figure
8-178.
Power Shut Down on page 4 -
S
M
ASIC
ERVICE
ANUAL
7.
8-147