P R E L I M I N A R Y
DIRECTION 5854713-100, REVISION 2
8-2-2

Battery Replacement Procedure

8-2-2-1

Tools

N/A
8-2-2-2

Time Required

5 min
8-2-2-3

Preparations

1.) Shut down the Venue Fit™ ultrasound unit, as described in
2.) Prepare stable flat surface covered with a soft material (such as foam or thick cloth) to prevent
scratches to the display.
3.) Dismount Venue Fit™ system from cart as described in
and place it on a previously prepared surface with the display facing down.
4.) Open the battery switch silicon cover and switch off the battery power.
Figure 8-30 Battery Power Switch
Chapter 8 - Replacement Procedures