8-5-9-5
ECG/Respiratory Module Assembly Installation Procedure
1) Position the ECG/Respiratory Module assembly in the Front End door assembly as shown in
Figure 8-130
2) Follow the ECG/Respiratory Module Assembly removal procedure described above, in reverse order.
3) Close the Front End door assembly door and fasten the two cage door securing latches.
4) Refit the following covers: front, DVD, right side, left side
5.) Turn ON power to the system.
The system re-boots into Normal Scanning mode.
6.) Press Config and log on as Adm - see
4 -
7.) Select About, then activate the HwVersion tab.
8.) Verify that the correct ECG/Respiratory module Part # is displayed, as shown in
ECG/Respiratory
Module
8-108
Section 8-5 - Electronic Cage Components - Replacement Procedures
1
EVISION
and
Figure
8-131.
Front Cover Installation Procedure on page 8 - 19
DVD Cover Installation Procedure on page 8 - 12
Right Side Cover Installation Procedure on page 8 - 10
Air Inlet (Left Side) Cover Installation Procedure on page 8 - 8
8.
Figure 8-132 Hardware Version Verification
V
S60N/V
IVID
Logging On to the Vivid™ S60/Vivid™ S70 as "ADM" on page
S70N
B
S
IVID
ASIC
ERVICE
Figure
M
ANUAL
8-132.