Battery Installation Procedure

THE Vivid™ S60/Vivid™ S70 BATTERY HAS A SAFETY DEVICE.
WARNING
WARNING
DO NOT ATTEMPT TO DIS-ASSEMBLE OR ALTER THE BATTERY!
Always observe the precautions listed in
1 -
22.
USE ONLY BATTERIES APPROVED BY GE HEALTHCARE AS SUITABLE FOR USE WITH THE
CAUTION
Vivid™ S60/Vivid™ S70 ULTRASOUND SCANNER
1) Insert a replacement battery into the battery compartment as shown in
Figure 8-177 Installation of Battery in AC Distribution Box Compartment
2.) Make sure the battery is properly inserted all the way.
3) Return the two securing screws (previously removed) -
4) Refit the AC Distribution Box cover:
5.) Turn ON power to the Vivid™ S60/Vivid™ S70 ultrasound scanner.
8-146
Section 8-5 - Electronic Cage Components - Replacement Procedures
1
EVISION
AC Distribution Box Cover Removal Procedure on page 8 - 29
V
S60N/V
IVID
IVID
Vivid S60N/Vivid S70N External Labels on page
Figure
8-176.
S70N
B
S
M
ASIC
ERVICE
Figure 8-177
below.