P R E L I M I N A R Y
D
FR091521, R
IRECTION
10-5-4

Cleaning

10-5-4-1

General Cleaning

Frequent and diligent cleaning of the Vivid™ S60/Vivid™ S70 ultrasound unit reduces the risk of
spreading infection from person to person, and also helps to maintain a clean working environment.
WHEN PERFORMING CLEANING PROCEDURES, TO PREVENT THE RISK OF SYSTEM
CAUTION
DAMAGE, ALWAYS OBSERVE THE FOLLOWING PRECAUTIONS:
Use only cleaning materials and solutions as recommended in the procedures described in the
Vivid™ S60/Vivid™ S70 User Manual.
Do not use any solutions or products not listed in the Vivid™ S60/Vivid™ S70 User Manual.
Never use thinner, benzene, ethanol or methanol alcohol, abrasive cleaners, or other strong
solvents, as these may cause damage to the cabinet or LCD panel. Only use isopropyl alcohol,
when instructed to do so.
Do not spray any liquid directly onto the Vivid™ S60/Vivid™ S70 covers, LCD Display or keyboard!
Do not allow any liquid to drip or seep into the system.
DO NOT scratch or press on the panel with any sharp objects, such as pencils or pens, as this may
result in damage to the panel.
Make sure not to spill or spray any liquid on the controls, into theVivid™ S60/Vivid™ S70 cabinet,
or in the probe connection receptacle.
Prior to cleaning, turn OFF power to the Vivid™ S60/Vivid™ S70 and disconnect the mains cable.
NOTE:
Refer to the Vivid™ S60/Vivid™ S70 User Manual for cleaning instructions