Touch Screen Rotaries Board Replacement Procedure; Tools; Time Required; Preparation - GE VIVID S60N Basic Service Manual

Hide thumbs Also See for VIVID S60N:
Table of Contents

Advertisement

P R E L I M I N A R Y
D
FR091521, R
IRECTION
8-3-5

Touch Screen Rotaries Board Replacement Procedure

8-3-5-1

Tools

Phillips screwdriver.
FRU Part #
Refer to
8-3-5-2

Time Required

30 min
8-3-5-3

Preparation

Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
8-3-5-4

Touch Screen Rotaries Board Removal Procedure

1) Raise the console to the maximum up position.
Make sure it is aligned in the central position (not pulled to one side or the other).
2) Remove the following Touch Screen covers: service, rear:
3) Remove the Touch Screen:
4.) Remove the six rotary knobs from the Rotaries Board -
NOTE:
After removing the Touch Screen Rotary Knobs, keep them safely, together with the screws, as they
will be required for refitting to the new Touch Screen Rotaries Board.
5.) Carefully lay the Touch Screen face-down on a clean, soft, stable surface.
6.) Remove the 8 securing screws from the Touch Screen Rotaries Board -
8-42
Section 8-3 - Control Console Components - Replacement Procedures
1
EVISION
Table 9-11
on page 9-10
Touch Screen Service Cover Removal Procedure on page 8 - 34
Touch Screen Rear Cover Removal Procedure on page 8 - 36
Touch Screen Removal Procedure on page 8 - 38
Figure 8-47 Removing the Touch Screen Rotaries Board
.
V
S60N/V
S70N
IVID
IVID
Power Shut Down on page 4 -
Figure
8-48.
Figure
B
S
M
ASIC
ERVICE
ANUAL
7.
8-47.