Right Side Cover Replacement Procedure; Tools; Time Required; Preparations - GE VIVID S60N Basic Service Manual

Hide thumbs Also See for VIVID S60N:
Table of Contents

Advertisement

P R E L I M I N A R Y
D
FR091521, R
IRECTION
8-2-3

Right Side Cover Replacement Procedure

8-2-3-1

Tools

None
FRU Part #
Refer to
8-2-3-2

Time Required

5 min
8-2-3-3

Preparations

Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
8-2-3-4

Right Side Cover Removal Procedure

Note:
The right side cover is secured to the mechanical platform by 4 snap-lock securing clips on
the inner side of the cover - illustrated below.
1) Gripping the recess at the bottom of the cover pull the right side cover out towards you to release
it from the snap-lock securing clips
2) Pull the top of the cover out towards you, releasing it from the upper securing clips.
3) Remove the cover from the system.
8-2-3-5

Right Side Cover Installation Procedure

1) Return the right side cover to the system, carefully aligning it with the securing clips.
2) Holding the recess with one hand, push the top of the cover upwards and inwards with the other
hand, until clicks into place.
3) Push the bottom of the cover upwards and inwards, until it clicks into place.
8-10
1
EVISION
Table 9-8
.
on page 9-7
Securing Clips
Figure 8-5 Removing the Right Side Cover
Section 8-2 - Covers - Replacement Procedures
V
S60N/V
IVID
(Figure
8-5).
S70N
B
S
IVID
ASIC
ERVICE
Power Shut Down on page 4 -
M
ANUAL
7.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivid s70n

Table of Contents