Manual Release Test Procedure; General; Test Equipment; Test Conditions - GE DINAMAP Compact TS Service Manual

Vital signs monitor
Table of Contents

Advertisement

4

Manual Release Test Procedure

The following procedure can be used to check the primary operation of the DINAMAP
order to release the unit for clinical use.

4.1 General

The DINAMAP
Compact unit is to be tested by applying various stimuli to the units' interfaces or
measuring unit parameters. The unit test procedure utilizes functions from the operational software and also
a service test mode attached to the operational software. A guide to the UUT Rear Connector and Front
Panel layout can be found in section 9.

4.1.1 Test Equipment

a.) Pressure Gauge 0-375mmHg, 0.2% accuracy, Digitron P200L or similar
b.) Nellcor Puritan Bennett™ SpO
c.) NIBP calibration kit part number 320-685
d.) Temperature Calibration Plug part number 320-635

4.1.2 Test Conditions

a.) Testing shall be conducted with an ambient temperature of 25 ? C ? 5 ? C

4.1.3 Service mode

4.1.3.1
Service Mode Selection
During the test procedure the operator will be requested to enter the Service Mode.
The UUT can be switched into its Service Mode by using the following procedure.
From the main menu select MODE
From the sub menu select SERVICE.
Enter the following code 2 2 1 3.
The UUT should then display the Service Menu.
4.1.3.2
Super Service Test Mode Selection
During the test procedure the operator will be requested to enter the Super Service Test Mode.
The UUT can be switched into its Super Service Test mode by using the following procedure:
From the main menu select MODE
From the sub menu select SERVICE.
Enter the following code 8 3 7 8.
The UUT should then display the Super Service Test Menu.
Simulator Model SRC-2, PT2500 or N1290 or similar
2
Compact monitor in
51

Advertisement

Table of Contents
loading

Table of Contents