Fan (TRX32 Box) Replacement Procedure

CAUTION
8-5-17-1

Tools

Phillips screwdriver.
FRU Part #
Refer to
8-5-17-2

Time Required

15 minutes
8-5-17-3

Preparation

Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
8-5-17-4

Fan (TRX32 Box) Removal Procedure

1) Remove the following covers: left side, right side, DVD, front.
2) Open and remove the Front End door assembly and place it on a stable surface:
3) Loosen the 2 screws that secure metal cover on the TRX32 Box and open the cover, as shown in
Figure 8-153 on page
8-130
Section 8-5 - Electronic Cage Components - Replacement Procedures
1
EVISION
When performing these procedures, take precautions to avoid damage of
electrostatic-sensitive components. Always have the ESD wrist strap
connected either to the DIB chassis or to the GND plug at the rear of the
scanner, and to your hand.
If a battery is present, first remove the battery as it contains stored energy.
Refer to
Battery Removal Procedure on page 8 -
Table 9-17
on page 9-16
Air Inlet (Left Side) Cover Removal Procedure on page 8 - 7
Right Side Cover Removal Procedure on page 8 - 10
DVD Cover Removal Procedure on page 8 - 11
Front Cover Removal Procedure on page 8 - 16
Front End Door Assembly Opening and Removal Procedure on page 8 - 83