8-4-3
Flex HDR (CFE to BEP) Cable Replacement Procedure
8-4-3-1
Tools
Phillips screwdriver.
FRU Part #
Table 9-20
Refer to
8-4-3-2
Time Required
10 min
8-4-3-3
Preparation
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
8-4-3-4
Flex HDR (CFE to BEP) Cable Removal Procedure
1) Remove the following covers: left side, right side, DVD, front.
Air Inlet (Left Side) Cover Removal Procedure on page 8 - 7
Right Side Cover Removal Procedure on page 8 - 10
DVD Cover Removal Procedure on page 8 - 11
Front Cover Removal Procedure on page 8 - 16
2) Open the Front End door assembly and secure the cage door using the support arm as described
in the
3.) Unscrew the two securing screws securing the HDR flex cable bracket located on the BEP as
shown in
4.) Disconnect the HDR flex cable from the Cage as described in the
Opening and Removal Procedure on page 8 -
1
EVISION
.
on page 9-19
Front End Door Assembly Opening and Removal Procedure on page 8 -
Figure 8-102
below.
Figure 8-102 HDR Flex Cable Bracket Located on BEP
Chapter 8 - Replacement Procedures
V
S60N/V
S70N
IVID
IVID
Power Shut Down on page 4 -
Front End Door Assembly
83.
B
S
M
ASIC
ERVICE
ANUAL
7.
83.
8-81