Probe Cable Holder Replacement Procedure; Time Required; Probe Cable Holder Removal Procedure; Probe Cable Holder Installation Procedure - GE VIVID S60N Basic Service Manual

Hide thumbs Also See for VIVID S60N:
Table of Contents

Advertisement

P R E L I M I N A R Y
D
FR091521, R
IRECTION
8-6-12

Probe Cable Holder Replacement Procedure

8-6-12-1
Tools
Use the appropriate Allen key as indicated in the Probe Cable Holder replacement procedure.
FRU Part #
Table 9-6
Refer to
8-6-12-2

Time Required

30 min
8-6-12-3
Preparations
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
8-6-12-4

Probe Cable Holder Removal Procedure

1) Raise the console to the maximum up position.
2) Working from the front of the system, use an Allen key to unscrew and remove the two screws that
fasten the Cable Holder bracket to the Support Arm, as shown in
Figure 8-202 Removing the Cable Holder Bracket from the Support Arm
3.) Remove the Cable Holder Bracket.
8-6-12-5

Probe Cable Holder Installation Procedure

1) Working from the front of the system, place the replacement Cable Holder bracket into position on
the Support Arm.
2) Return the two screws (previously removed) and use an Allen key to fasten the screws to secure
the Cable Holder firmly in place - refer to
3) Proceed to route the probe cables into the correct position, as described in
page
1
EVISION
.
on page 9-5
8-168.
Chapter 8 - Replacement Procedures
V
S60N/V
IVID
IVID
Figure
8-202.
S70N
B
S
M
ASIC
ERVICE
Power Shut Down on page 4 -
Figure
8-202.
section 8-6-12-5-1 on
ANUAL
7.
8-167