Vivid™ S60/Vivid™ S70 Touch Screen Replacement Procedure; Tools; Time Required; Preparation - GE VIVID S60N Basic Service Manual

Hide thumbs Also See for VIVID S60N:
Table of Contents

Advertisement

P R E L I M I N A R Y
D
FR091521, R
IRECTION
8-3-4
Vivid™ S60/Vivid™ S70 Touch Screen Replacement Procedure
NOTE:
The procedures below provide instructions for replacement of the Touch Screen.
For details on replacing the Touch Screen chassis, refer to
Procedure on page 8 -
8-3-4-1

Tools

Phillips screwdriver and 7mm socket.
FRU Part #
Refer to
NOTE:
When replacing the Touch Screen, if replacement of the Touch Screen Rotaries Board is also required,
it is also necessary to order this part (supplied separately). See
8-3-4-2

Time Required

20 min
8-3-4-3

Preparation

Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
8-3-4-4

Touch Screen Removal Procedure

1) Raise the console to the maximum up position.
Make sure it is aligned in the central position (not pulled to one side or the other).
2) Remove the following Touch Screen covers: service, rear:
Note:
The Touch Screen Panel is fastened to the Touch Screen chassis by 4 screws (2 on each side).
3) Remove the 2 securing screws on both the left and right sides of the Touch Screen -
8-38
Section 8-3 - Control Console Components - Replacement Procedures
1
EVISION
147.
Table 9-11
on page 9-10 (Touch Panel Assembly)
Touch Screen Service Cover Removal Procedure on page 8 - 34
Touch Screen Rear Cover Removal Procedure on page 8 - 36
Figure 8-40 Removing the Touch Screen Securing Screws
V
S60N/V
S70N
IVID
IVID
Touch Screen Chassis Replacement
Table 9-11
on page 9-10.
Power Shut Down on page 4 -
B
S
M
ASIC
ERVICE
ANUAL
7.
Figure