GE H
EALTHCARE
D
FL091021, R
IRECTION
8-2-2
Hard Disk Replacement Procedure
8-2-2-1
Tools
Use the appropriate flat and Phillips-type screw drivers as indicated in the Hard Disk replacement procedure.
8-2-2-2
Time Required
10 minutes
8-2-2-3
Preparation
Shut down the Vivid i n ultrasound unit, as described in
8-2-2-4
Hard Disk Removal Procedure
1) Make sure the Vivid i n LCD Display cover is securely closed and that the scanner is standing firmly
on a flat surface.
2) Working from the rear of the scanner (rear panel), use a small Phillips screwdriver to loosen and
remove the two screws that secure the Hard Disk in the hard disk module holder, as shown in
Figure 8-3
3) With both hands, tilt the rear end of the scanner upwards then, holding it steady with one hand,
remove the Hard Disk from the system by holding the grip and pulling it straight out of the hard disk
module holder, as shown in
.
8-4
5
EVISION
below.
Figure 8-3 Removing the Screws from the Rear Panel
Figure 8-4 Removing the Hard Disk
Section 8-2 - External Component Replacement Procedures
Figure 8-4
below.
V
IVID I N
Chapter
3-
System
Setup.
S
M
ERVICE
ANUAL