Hard Disk Replacement Procedure
WHILE THE SOFTWARE INSTALL PROCEDURE IS DESIGNED TO PRESERVE DATA,
WARNING
WARNING
YOU SHOULD SAVE ANY PATIENT DATA, IMAGES, SYSTEM SETUPS TO A DVD OR
HARDCOPY BEFORE DOING A SOFTWARE UPGRADE.
NOTE:
When replacing the BEP or Hard Disk, one must either obtain new access codes that will fit the
system ID which is programmed on the BEP, or program the BEP system ID to fit the access codes
already registered in the Hard Disk. Only qualified personnel should perform these tasks.
8-5-3-1

Tools

Appropriate flat and Phillips-type screwdrivers, and Hard Disk Duplicator.
FRU Part #
Table 9-14
Refer to
8-5-3-2

Time Required

10 minutes
8-5-3-3

Preparation

Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
8-5-3-4

Hard Disk Removal Procedure

1) Lock the wheels, then remove the following covers: left side, right side, DVD, front.
Air Inlet (Left Side) Cover Removal Procedure on page 8 - 7
Right Side Cover Removal Procedure on page 8 - 10
DVD Cover Removal Procedure on page 8 - 11
Front Cover Removal Procedure on page 8 - 16
Note:
A metal plate holder secures the hard disk as shown in
secure the hard disk firmly in position on the Back End and also serves as a useful hand
grip when removing/inserting the hard disk, as described in the following steps.
Metal Securing Plate
attached to Hard Disk
Hard Disk retaining
screw
1
EVISION
.
on page 9-13
Metal Securing Plate
Figure 8-114 Hard Disk with Metal Securing Plate Attached
Chapter 8 - Replacement Procedures
V
S60N/V
S70N
B
IVID
IVID
Power Shut Down on page 4 -
Figure
8-114. The plate is used to
S
M
ASIC
ERVICE
ANUAL
7.
Hard Disk
8-91

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivid s70n

Table of Contents