P R E L I M I N A R Y
D
5813707-100, R
IRECTION
8-2-2

Batteries Replacement Procedure

8-2-2-1
Tools
N/A
8-2-2-2
Time Required
5 min
8-2-2-3
Preparations
1.) Shut down the Venue Go™ ultrasound unit, as described in
2.) Prepare stable flat surface covered with a soft material (such as foam or thick cloth) to prevent
scratches to the display.
3.) Dismount Venue Go™ system from cart as described in
and place it on a previously prepared surface with the display facing down.
4.) Open the battery switch silicon cover and switch off the battery power.
2
EVISION
Figure 8-71 Battery Power Switch
Chapter 8 - Replacement Procedures
V
™ S
ENUE
Power
Shutdown.