Physical Inspection - GE VIVID S60N Basic Service Manual

Hide thumbs Also See for VIVID S60N:
Table of Contents

Advertisement

P R E L I M I N A R Y
D
FR091521, R
IRECTION
10-5-3

Physical Inspection

Table 10-8
Step
Item
1
Labeling
2
Scratches & Dents
3
Covers
4
Input Power
5
External I/O
6
Wheels and Brakes
7
Control Panel Movement
8
Control Panel Lighting
9
LCD
10
Monitor Light
11
Cables and Connectors
12
Shielding and Covers
13
Control Panel
14
Probe Holders
Power and System
15
Status Indicators
16
Battery
1
EVISION
Physical Checks
Verify that all Ultrasound system labeling is present and in readable condition.
Inspect the exterior for dents, scratches or cracks.
Where applicable, verify all covers are secured in place and are properly aligned with other covers.
Replace any covers that are damaged.
Refer to:
Mains Cable Inspection on page 10 -
Check all connectors for damage.
• Where applicable, check all wheels and casters for wear and verify operation of foot brake, to stop
the Ultrasound system from moving, and release mechanism.
• Where applicable, check all wheel locks and wheel swiveling for proper operation.
• Where applicable, verify ease of Operator Panel (Operator Control Panel) movement in all
acceptable directions.
• Where applicable, ensure that the Control panel latches in position as required.
Check for proper operation of all operator panel and TGC lights.
Inspect the LCD Display for scratches and bad pixels.
Verify proper operation of Contrast and Brightness controls.
Where applicable, confirm that the LCD arm allows:
• Swiveling the screen to the left and to the right
• Folding the screen to the locked position
• Release and adjustment backwards and forwards
• Can be adjusted in the up/down positions.
Note: LCD Arm movement may vary and is not applicable to all Ultrasound systems.
Check for proper operation of any monitor lighting, if available.
Check all internal cable harnesses and connectors for wear and secure connector seating. Pay
special attention to probe strain or bend reliefs.
Check to ensure that all EMI shielding, internal covers, air flow panels and screws are in place.
Missing covers and hardware could cause EMI/RFI problems while scanning.
Inspect alphanumeric keyboard and Operator Panel. Record any damaged or missing items.
Where applicable, inspect the Probe Holders for cracks or damage.
Check for proper operation of all Power and System Status Indicators.
Where applicable, check that the battery is not damaged, does not leak, does not emit an odor, and
is not deformed or discolored. Observe all warnings and cautions for battery handling, recharging,
storing, and/or disposal.
Chapter 10 - Care and Maintenance