Optimizing M-Mode
The use of preset gives optimum performance with minimum adjustment. If necessary, the following
controls can be adjusted to further optimize the M-Mode display:
•
Adjust Horizontal sweep to optimize the display resolution.
•
Adjust Gain and TGC controls to adjust the range to be imaged.
•
Use the Frequency (move to higher frequencies) or the Frame rate control (move to lower frame
rate) to increase resolution in image.
•
Adjust Dynamic range to optimize the useful range of incoming echoes to the available greyscale.
•
Adjust Compress to further optimize the display.
•
Adjust Reject to reduce noise while taking care not to eliminate significant low-level diagnostic
information.
4-3-5
Color Mode Checks
4-3-5-1
Introduction
Color Flow screens are 2D or M Mode screens with colors representing blood or tissue movement.
Color Flow may be selected both from 2D mode or from M mode or a combination of these.
4-3-5-2
Color 2D Mode Overview
1. Probe orientation marker
2. Color bar
3. Color sector marker
4. Parameter window
1
EVISION
Figure 4-16 Color Mode Screen