Fan (AC Distribution Box) Replacement Procedure

8-5-22-1

Tools

Phillips screwdriver.
FRU Part #
Refer to
8-5-22-2

Preparation

Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
8-5-22-3

Time Required

10 minutes
8-5-22-4

Fan (AC Distribution Box) Removal Procedure

NOTE:
Access to the Fan is from the right side of the AC Distribution Box -
1.) Remove the AC Distribution Box and place it on a clean stable surface:
8-142
Section 8-5 - Electronic Cage Components - Replacement Procedures
1
EVISION
Table 9-17
on page 9-16
Figure 8-172 AC Distribution Box Fan - Location
AC Distribution Box Removal Procedure on page 8 - 139
.