Fan for Cabinet Cage Assembly Replacement Procedure
8-5-19-1
Tools
Phillips screwdriver.
FRU Part #
Refer to
8-5-19-2
Time Required
20 mins
8-5-19-3
Preparation
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
8-5-19-4
Fan for Cabinet Cage Assembly Removal Procedure
The fan (see
1) Remove the following covers: right side, DVD, front, right rear:
•
•
•
•
2.) Disconnect the Fan cable as indicated in
Disconnect cable
Securing screw
3.) Loosen and remove the two securing screws as indicated in
4.) Remove the fan.
8-136
Section 8-5 - Electronic Cage Components - Replacement Procedures
1
EVISION
Table 9-17
on page 9-16
Figure 8-165
below) is located at the rear of the scanner.
Right Side Cover Removal Procedure on page 8 - 10
DVD Cover Removal Procedure on page 8 - 11
Front Cover Removal Procedure on page 8 - 16
Right Rear Cover Removal Procedure on page 8-20
Figure 8-165 Fan for Cabinet Cage Assembly
.
Figure 8-165
V
S60N/V
S70N
IVID
IVID
Power Shut Down on page 4 -
below.
Securing screw
Figure 8-165,