Cabinet Cage Module Replacement Procedure
8-5-2-1
Tools
Phillips screwdriver and a 3mm Allen key.
FRU Part #
Refer to
8-5-2-2
Time Required
45 mins
8-5-2-3
Preparation
Shut down the Vivid S60/Vivid S70 ultrasound unit, as described in
8-5-2-4
Cabinet Cage Module Removal Procedure
1) Remove the following covers: left side, right side, DVD and Printer, front, right rear, left rear:
2) Remove the Front End Door Assembly, BEP, and Sub-Woofer Assembly:
3.) Disconnect all cables from the Internal BIF Panel located above the Cabinet Cage Module - see
Figure
8-88
Section 8-5 - Electronic Cage Components - Replacement Procedures
1
EVISION
Table 9-6
.
on page 9-5
Air Inlet (Left Side) Cover Removal Procedure on page 8 - 7
Right Side Cover Removal Procedure on page 8 - 10
DVD and Printer Cover Removal Procedure on page 8 - 13
Front Cover Removal Procedure on page 8 - 16
Right Rear Cover Removal Procedure on page 8 - 20
Left Rear Cover Removal Procedure on page 8 - 23
Front End Door Assembly Opening and Removal Procedure on page 8 - 83
BEP Removal Procedure on page 8 - 94
Sub-Woofer Assembly Removal Procedure on page 8 - 133
8-112.
Figure 8-112 Disconnect Cables from Internal BIF Panel