Peripheral/Option Checks; Mains Cable Inspection - GE VIVID S60N Basic Service Manual

Hide thumbs Also See for VIVID S60N:
Table of Contents

Advertisement

P R E L I M I N A R Y
D
FR091521, R
IRECTION
10-5-2-2

Peripheral/Option Checks

If any peripherals or options are not part of the system configuration, the check can be omitted.
Refer to the Vivid™ S60/Vivid™ S70 User Manual for a list of approved peripherals/options.
Table 10-6
Step
1
2
3
4
5
6
10-5-2-3

Mains Cable Inspection

Table 10-7
Step
1
2
3
4
10-8
1
EVISION
GE Approved Peripheral/Hardware Option Functional Checks
Item
Media
Verify media drive(s) read/write properly. Clean if necessary.
B/W Printer
Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
Color Printer
Verify hardcopy output of the Color video page printer. Clean heads and covers if necessary.
DICOM
Verify that DICOM is functioning properly. Send an image to a DICOM device.
ECG
Verify basic operation with customer.
Footswitch
Verify that the footswitch is functioning as programed. Clean as necessary.
Mains Cable Inspection, As Appropriate
Item
Unplug Cord
Disconnect the mains cable from the wall outlet and from the Ultrasound system.
Mains Cable
Inspect the mains cable and its connectors for any damage.
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals,
Terminals
and that no strands may cause a short circuit.
Inlet Connector
Verify that the Inlet connector retainer is functional.
Section 10-3 - Why Perform Maintenance Procedures?
V
S60N/V
S70N
IVID
IVID
Description
Description
B
S
M
ASIC
ERVICE
ANUAL

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivid s70n

Table of Contents