Isolated Patient Lead (Source) Leakage - Lead-To-Lead; Lead-To-Lead Leakage Test Record - GE VIVID S60N Basic Service Manual

Hide thumbs Also See for VIVID S60N:
Table of Contents

Advertisement

P R E L I M I N A R Y
D
FR091521, R
IRECTION
10-6-8
Isolated Patient Lead (Source) Leakage – Lead-to-Lead
When using any calibrated Electrical Safety Analyzer tool compliant with AAMI/ESI 1993 or IEC 60601
or AS/NZS 3551, switch the meter's function selector to the LEAD-LEAD position (or equivalent function
in the device).
Select and test each of the ECG lead positions (except ALL) on the LEAD selector, testing each to the
power and ground condition combinations found in:
Record the highest leakage current measured.
10-6-8-1

Lead-to-Lead Leakage Test Record

Table 10-16
Measurements should be recorded from each lead combination under each set of test conditions
specified in:
Record all data on the EQC inspection certificate; also known as Patient Auxiliary Current.
1) Connect Safety analyzer to wall AC power outlet.
2) Plug the equipment under the test power cable into the receptacle on the meter's panel.
3) Connect the ECG cable to the Vivid™ S60/Vivid™ S70 ultrasound system and the Patient leads to
the analyzer.
4) Select the Patient lead leakage function on the meter.
5) Test opening and closing the ground with the Vivid™ S60/Vivid™ S70 ultrasound system ON and
OFF, as indicated in
NOTE:
Refer to the safety analyzer's user manual that will be used to perform the tests.
Table 10-16 Typical Data Format for Recording Patient Lead-to-Lead Leakage
Unit under test____________________________________
System Power
System Off
System On
(Transmit)
10-24
1
EVISION
Table 10-11
on page 10-17
or
Table 10-12
on page 10-17
as applicable.
below shows a typical format for recording the patient lead-to-lead leakage current.
Table 10-11
on page 10-17
or
Table 10-12
on page 10-17
as applicable.
Table 10-11
Test Conditions
Grounding/PE
open
open
Section 10-3 - Why Perform Maintenance Procedures?
V
S60N/V
IVID
on page 10-17 or
Table 10-12
Date of test:_____________
Patient Lead or Combination Measured
RA-LA
LA-LL
S70N
B
S
IVID
ASIC
ERVICE
on page 10-17, as applicable.
LL-RA
M
ANUAL