10-5-2

Functional Checks

NOTE:
Refer also to Chapter 4 -General Procedures and Functional Checks, for additional details about the
functional checks described in this section.
The functional checks take approximately 60 minutes to perform. Refer to the Vivid™ S60/Vivid™ S70
User Manual whenever necessary.
10-5-2-1

System Checks

Step
Item (or Mode)
1
B-Mode
2
CF-Mode
3
Doppler Modes
4
M-Mode
5
Probe Elements
6
Applicable Software Options
8
Operator Panel test
9
Keyboard
10
Touch Panel
11
Monitor
12
Peripherals
1
Table 10-5
System Functional Checks
Verify basic B-Mode (2D) operation.
Check the basic Ultrasound system controls that affect this mode of operation.
Verify basic CF-Mode (Color Flow Mode) operation.
Check the basic Ultrasound system controls that affect this mode of operation.
Verify basic Doppler operation (PW and CW if available).
Check the basic Ultrasound system controls that affect this mode of operation.
Verify basic M-Mode operation.
Check the basic Ultrasound system controls that affect this mode of operation.
Perform an Element Test on each probe to verify that all the probe elements and system
channels are functional.
Verify the basic operation of all optional modes such as Contrast. Check the basic
Ultrasound system controls that affect each option's operation.
Perform the Operator Panel Test Procedure.
Do the interactive keyboard test.
Verify basic Touch Panel display functions.
Verify basic monitor display functions.
See:
Chapter 10 - Care and Maintenance
V
S60N/V
IVID
Peripheral/Option Checks on page 10 -
S70N
B
S
IVID
ASIC
Description
8.
M
ERVICE
ANUAL
10-7