Section 8-5
Electronic Cage Components - Replacement Procedures
NOTE:
When replacing Vivid™ S60/Vivid™ S70 lower section components, after removal of the system covers (left
and right sides; front), the Front End door assembly should be removed from the main assembly (to gain
access to the internal components) and placed on a stable surface. These steps are described in the procedure
below.
CAUTION
8-5-1
Opening and Removing the Front End Door Assembly
8-5-1-1
Tools
Appropriate Phillips screwdriver
8-5-1-2
Time Required
5 min
8-5-1-3
Preparation
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
8-5-1-4
Front End Door Assembly Opening and Removal Procedure
1) Remove the following covers: left side, right side, DVD, front.
•
Air Inlet (Left Side) Cover Removal Procedure on page 8 - 7
•
Right Side Cover Removal Procedure on page 8 - 10
•
DVD Cover Removal Procedure on page 8 - 11
•
Front Cover Removal Procedure on page 8 - 16
2) Open the two fastening latches on the left side of the Front End door assembly and carefully open
the cage door.
Figure 8-104 Unfastening the Latches and Opening the Cage Door
1
When performing these procedures, take precautions to avoid damage of
electrostatic-sensitive components. Always have the ESD wrist strap
connected either to the DIB chassis or to the GND plug at the rear of the
scanner, and to your hand.
If a battery is present, first remove the battery as it contains stored energy.
Refer to
Battery Removal Procedure on page 8 -
(Figure
8-104).
Chapter 8 - Replacement Procedures
V
S60N/V
S70N
B
IVID
IVID
145.
Power Shut Down on page 4 -
S
M
ASIC
ERVICE
ANUAL
7.
8-83