Ecg/Respiratory Module Assembly Replacement Procedure; Tools; Time Required; Preparation - GE VIVID S60N Basic Service Manual

Hide thumbs Also See for VIVID S60N:
Table of Contents

Advertisement

P R E L I M I N A R Y
D
FR091521, R
IRECTION
8-5-9

ECG/Respiratory Module Assembly Replacement Procedure

CAUTION
8-5-9-1

Tools

Appropriate Phillips screwdriver.
FRU Part #
Refer to
8-5-9-2

Time Required

10 mins
8-5-9-3

Preparation

Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
8-5-9-4

ECG/Respiratory Module Assembly Removal Procedure

1) Remove the following covers: left side, right side, DVD, front.
2) Open the Front End door assembly, as follows:
a.) Open the two fastening latches on the side of the Front End door assembly and carefully open
b.) Secure the cage door with the cage door securing arm - see
NOTE:
The location of the ECG/Respiratory Module Assembly is shown in
8-106
Section 8-5 - Electronic Cage Components - Replacement Procedures
1
EVISION
When performing these procedures, take precautions to avoid damage of
electrostatic-sensitive components. Always have the ESD wrist strap
connected either to the DIB chassis or to the GND plug at the rear of the
scanner, and to your hand.
If a battery is present, first remove the battery as it contains stored energy.
Refer to
Battery Removal Procedure on page 8 -
Table 9-18
on page 9-17
Air Inlet (Left Side) Cover Removal Procedure on page 8 - 7
Right Side Cover Removal Procedure on page 8 - 10
DVD Cover Removal Procedure on page 8 - 11
Front Cover Removal Procedure on page 8 - 16
the cage door.
.
V
S60N/V
S70N
IVID
IVID
145.
Power Shut Down on page 4 -
Figure 8-105 on page
Figure
8-130.
B
S
M
ASIC
ERVICE
ANUAL
7.
8-84.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivid s70n

Table of Contents