3-4-3-4

Peripheral/Accessory Connector Panel

Figure 3-18
peripheral/accessory connectors.
5
6
Figure 3-18 View of the Vivid S60N/Vivid S70N Peripheral/Accessory Connector Panel
1
Ethernet LAN connector — 1000 Base-TX Ethernet IEEE 802.3
2
Isolated USB connector (USB 1.0 only)
3
Dual USB connector
4
DVI-D Display OUT connector (DVI-I type with digital output only [DVDI-D])
5
LED - Network activity
6
LED - Network activity
3-4-4

EMI Protection

The Vivid™ S60/Vivid™ S70 has been designed to minimize the effects of Electro-Magnetic
Interference (EMI). Many of the covers, shields, and screws are provided primarily to protect the Vivid™
S60/Vivid™ S70 from image artifacts caused by this interference. For this reason, it is imperative that
all covers and hardware are installed and secured before the Vivid™ S60/Vivid™ S70 is put into
operation.
See
EMI Limitations on page 2 - 4
3-18
1
EVISION
shows a view of the Vivid S60N/Vivid S70N ultrasound unit rear panel showing external
1
Section 3-3 - Receiving and Unpacking the Equipment
V
IVID
2
3
for more information about EMI protection.