Battery - General Safety Guidelines - GE VIVID S60N Basic Service Manual

Hide thumbs Also See for VIVID S60N:
Table of Contents

Advertisement

P R E L I M I N A R Y
D
FR091521, R
IRECTION
5-11-3-1

Battery - General Safety Guidelines

The lithium ion rechargeable battery provides a backup mechanism to the Vivid™ S60/Vivid™ S70
system whenever an AC power source is not available. The battery module is supplied with a lithium
ion rechargeable battery pack (GPA) installed in the battery bay, as standard.
The Vivid™ S60/Vivid™ S70 has built-in charger functionality and switches automatically from battery
operation to AC operation and vice versa.
When shutting down the system, leave the main power cable connected to keep the battery fully charged.
NOTE:
Before removing or inserting the Battery, perform system shut-down and disconnect the AC power
cable from the Vivid™ S60/Vivid™ S70 ultrasound scanner.
NOTE:
The lithium ion technology used in the system's battery is significantly less hazardous to the
environment than the lithium metal technology used in some other batteries.
NOTE:
The battery is designed to be replaced every 2 years.
THE BATTERY IS DESIGNED TO WORK WITH VIVID™ S60/VIVID™ S70 SYSTEMS ONLY.
CAUTION
ONLY USE THE BATTERIES AUTHORIZED BY GE.
-
-
-
-
-
-
-
-
-
-
-
-
-
Recommended storage conditions of battery pack:
Short term (less than one month):
Long term (more than three months): 10 °C (50 °F) to 35 °C (95 °F).
1
EVISION
Do not disassemble or alter it. Charge the batteries only when the ambient temperature is
between 0 °C and 65 °C (32 °F and 149 °F) and discharge the batteries between -10 °C and
55 °C (14 °F and 131 °F).
Do not short-circuit the battery by directly connecting the battery terminals with metal objects.
Do not heat the battery or incinerate.
Do not expose the battery to temperature over 60 °C (140 °F). Keep it away from fire and
other heat sources.
Do not charge the battery near a heat source, e.g. fire or heaters.
Do not leave the battery in direct sunlight.
Do not pierce the battery with a sharp object, hit it, or step on it.
Do not use a damaged battery. Do not solder a battery.
Do not connect the battery to an electrical outlet.
Do not immerse the battery in water or allow it to get wet.
Do not put the battery into a microwave oven or pressurized container. If the battery leaks or
emits an odor, remove it from all possible flammable sources.
If the battery emits an odor or heat, is deformed or discolored, or in a way appears abnormal
during use, recharging or storage, immediately remove it and stop using it.
If you have any questions about the battery, consult GE or your local representative.
Chapter 5 - Components and Function (Theory)
V
S60N/V
IVID
0 °C (32 °F)
to 50 °C (122 °F)
S70N
B
S
IVID
ASIC
ERVICE
M
ANUAL
5-27

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivid s70n

Table of Contents