Battery Replacement Procedure

8-5-23-1

Tools

Phillips screwdriver.
FRU Part #
Table 9-13
Refer to
8-5-23-2

Time Required

1 min
8-5-23-3

Preparations

Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
8-5-23-4

Battery Removal Procedure

1.) Remove the AC Distribution Box cover:
AC Distribution Box Cover Removal Procedure on page 8 - 29
2.) Remove the 2 securing screws from the battery compartment -
3) Slide the battery out of the AC Distribution Box -
WHEN REMOVING A DEFECTIVE BATTERY, ENSURE THAT IT IS DISPOSED OF IN
CAUTION
ACCORDANCE WITH LOCAL REGULATIONS.
ALTERNATIVELY, FORWARD IT TO GE HEALTHCARE FOR PROPER DISPOSAL.
1
EVISION
.
on page 9-12
Figure 8-176 Removing the Screw on the Battery Compartment
Chapter 8 - Replacement Procedures
V
S60N/V
S70N
IVID
IVID
Power Shut Down on page 4 -
Figure 8-176
Figure 8-176
(right).
B
S
M
ASIC
ERVICE
ANUAL
7.
(left).
8-145