Swivel and Up-Down Handle Replacement Procedure
8-6-7-1
Tools
Flat and Phillips screwdrivers as needed.
8-6-7-2
Time Required
60 min
8-6-7-3
Preparation
FRU Part #
Table 9-5
Refer to
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
8-6-7-4
Swivel and Up-Down Handle Removal Procedure
1) Remove the Operator Panel as described in the
Procedure on page 8 -
2) Remove the Probe Shelf as described in the
3.) Unscrew the protective metal plate (shown in
Swivel Cable
Up-down Cable
1
EVISION
(Swivel Mechanism).
on page 9-3
44.
Figure 8-189 Protective Metal Plate Rotated 90 Degrees
Chapter 8 - Replacement Procedures
V
S60N/V
S70N
IVID
IVID
Power Shut Down on page 4 -
Operator Panel Keyboard Assembly Removal
Probe Shelf Removal Procedure on page 8 -
Figure
8-189) and rotate it upwards 90 degrees.
B
S
M
ASIC
ERVICE
ANUAL
7.
166.
8-155