Swivel and Up-Down Handle Replacement Procedure

8-6-7-1

Tools

Flat and Phillips screwdrivers as needed.
8-6-7-2

Time Required

60 min
8-6-7-3

Preparation

FRU Part #
Table 9-5
Refer to
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
8-6-7-4

Swivel and Up-Down Handle Removal Procedure

1) Remove the Operator Panel as described in the
Procedure on page 8 -
2) Remove the Probe Shelf as described in the
3.) Unscrew the protective metal plate (shown in
Swivel Cable
Up-down Cable
1
EVISION
(Swivel Mechanism).
on page 9-3
44.
Figure 8-189 Protective Metal Plate Rotated 90 Degrees
Chapter 8 - Replacement Procedures
V
S60N/V
S70N
IVID
IVID
Power Shut Down on page 4 -
Operator Panel Keyboard Assembly Removal
Probe Shelf Removal Procedure on page 8 -
Figure
8-189) and rotate it upwards 90 degrees.