BIF (Back End Interface) Board Replacement Procedure

CAUTION
NOTE:
The Back End Interface Board (BIF Board) is located at the top of the electronic cabinet cage, in effect
forming the roof of the cage. See
8-5-15-1

Tools

Phillips screwdriver.
FRU Part #
Refer to
8-5-15-2

Time Required

30 minutes
8-5-15-3

Preparation

Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
8-122
Section 8-5 - Electronic Cage Components - Replacement Procedures
1
EVISION
When performing these procedures, take precautions to avoid damage of
electrostatic-sensitive components. Always have the ESD wrist strap
connected either to the DIB chassis or to the GND plug at the rear of the
scanner, and to your hand.
If a battery is present, first remove the battery as it contains stored energy.
Refer to
Battery Removal Procedure on page 8 -
View from Underside
View from Top
Figure 8-147 BIF Board and Location in Electronic Cage
Table 9-14
on page 9-13
Figure 8-147
below.
.
V
S60N/V
S70N
IVID
IVID
145.
Location of BIF Board
Power Shut Down on page 4 -