Rear Handle Replacement Procedure

NOTE:
When replacing a rear handle, make sure the rear handle kit has been ordered, the contents of which
are listed in
Table 8-3
R2421749
R2420760
RD9130410
2419072
2418493
8-6-8-1

Tools

Allen Key 2mm
FRU Part #
Refer to
8-6-8-2

Time Required

20 min
8-6-8-3

Preparation

Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
8-6-8-4

Rear Handle Removal Procedure

NOTE:
The rear handles are secured to the aluminum bars with a mounting screw and a small amount of glue.
The screw is first released, then a twisting of the handles is required to release the glue before sliding
the handle off the aluminum bar. See the instructions below.
1.) Usin
handle.
2.) Gently twist the handle to release the glue, then remove the handle from the aluminum bar.
3.) Repeat
8-158
Section 8-6 - Mechanical Platform Components - Replacement Procedures
1
EVISION
Table
8-3, below.
Rear Handle Kit - PN S2424091
P/N
Left handle
Right Handle
Socket Set Screw Flat Point M4, L10, DIN 913
Package (White Mantle 260X236X63)
Good for stock Label
Table 9-7
.
on page 9-6
the Socket Set Screw Flat Point, loosen and remove the mounting screw from one of the rear
g
step 1
and
step 2
Description
to remove the second handle.