Variable Voltage Source (Opt. 309) - Keysight Technologies FieldFox A Series User Manual

Table of Contents

Advertisement

System Settings
System Configuration
To learn the expected measurement accuracy when connected to a frequency
reference source, see the FieldFox specifications in
Sheet" on page
Reference OUT
The internal 10 MHz frequency reference is available from the Ref/Trig output
connector on the right-side panel. Use this signal as a common frequency
reference when making measurements using external equipment. See this
connector in

Variable Voltage Source (Opt. 309)

An internal DC source, available with Opt. 309, may be used to provide power
for external devices. The maximum power available is 8 watts.
The voltage is output through an SMB (m) output connector located on the
left-side panel. See the connector in
An optional SMB (f) to BNC (m) bias-tee power cable is available with Keysight
part number: N9910X-713.
The following status line is provided near the top of the FieldFox screen.
Figure 29-9
Voltage Source Status Line
1. Voltage source state: ON, OFF, TRIPPED.
2. Voltage setting from 1.00 V to 32.00 V with.01 V resolution.
3. Current: The amount of current (in Amperes) being drawn from the source.
4. Max Current: A conservative estimate of maximum current possible given
656
717.
"Top Panel" on page
IMPORTANT! Noise Figure mode uses DC variable voltage source (VVS) –
(Option 309) during the measurements. The voltage might be as high as
28V during the measurement. During the noise figure measurement,
always make sure there are no voltage sensitive devices connected to the
DC SMB output port. Refer to
– CPU2 Only)", on page
When you initially change the FieldFox mode to Noise Figure, the FieldFox
displays the following message for ten seconds:
"Ensure your Noise Source is connected to the VVS DC Output
It (VVS) will be driven up to +28V during the measurement".
the present voltage setting. The actual 'trip' level may be slightly different.
38.
Chapter 27, "Noise Figure Mode (Option 356
547.
"Left Side Panel" on page
Keysight N9927-90001 User's Guide
"Specifications/Data
40.