Calibration for NA, CAT, and VVM Modes
— For each DUT port:
— Select/Change Cal Type. The FieldFox always RECOMMENDS a Cal Type
— Press
No Connection
— Disconnect the DUT from the FieldFox.
— If a jumper cable or adapter is required to connect the DUT to the FieldFox,
— Do NOT connect anything at the cal reference plane where the DUT
— Press
Optionally Connect Load(s)
— Connect a Load standard at the point where the DUT will be connected and
126
— The connector choices that are offered depend on the frequency
range. For example, if you have selected a frequency range that
exceeds 18 GHz, the only connector options you will be presented
with in QuickCal are 'other 50 ohm' or 'other 75 ohm'. This is because
3.5 mm connectors, which go beyond 18 GHz, are not supported with
QuickCal.
— Select the connector type of the DUT using the
knob.
— Press
Change Gender
— Press Next... to make the selection.
— If present, press Finish.
that will quickly and accurately calibrate all of the displayed S-parameters.
The other Cal Types that are listed will NOT necessarily calibrate all of the
displayed S-parameters. Learn more about Cal Types in
on page
134.
— Using the
Select
and Finish.
Start Calibration
connect those components to the FieldFox test ports connectors. The
effects of those components will be measured and removed during the
calibration, and only the effects of the DUT will be displayed in the
measurement results. These should be high-quality components!
connects.
Measure
press
Measure
under the following conditions:
— When a jumper cable has too much loss. If a 'Cal Failed' error
appears and you skipped the Load measurement, then perform the
Cal again but include measuring the Load.
to change the gender of the DUT connector.
arrows or rotary knob, select a Cal Type, then press
Keysight N9927-90001 User's Guide