Power Sensor Settling; Sweep Settings - Keysight Technologies FieldFox A Series User Manual

Table of Contents

Advertisement

Frequency Offset Using Power Sensor (FOPS) – Option 208
FOPS Settings

Power Sensor Settling

The following two settings work together to allow you to choose the right
compromise between measurement speed and accuracy.
Each power sensor reading is considered "settled" when either:
— The difference between two consecutive readings are within the Tolerance
— Or the Max (number of) Readings has been made.
The readings that were taken are averaged together to become the displayed
reading.
How to choose Power Sensor settling
— Press BW 2
— Then Tolerance. Enter a value in dB between 0 and 10 using the numeric
— Then

Sweep Settings

These settings provide added control of the FOPS measurement.
How to make additional sweep settings
— Press Sweep 3
— Then choose from the following:
466
Src = 2 GHz to 3 GHz
Offset = 3.8 GHz
Rx = 5.8 GHz to 6.8 GHz
value.
keypad, the
arrows, or the rotary knob. Tolerance of 0 will never be
achieved, so the Max Readings will always be made. Press
Max
Readings. Enter a value between 1 and 20 using the numeric
keypad, the
arrows, or the rotary knob. Press
Power Setup
Source Enable
——ON
Turn ON the internal source (required for FOPS).
——OFF
Turn OFF the internal source
Src Nom Power
arbitrary value at the source CW frequency. Then enter a value using
the numeric keypad, the
(-15 dBm default setting) Set output power to an
arrows, or the rotary knob. Press
Keysight N9927-90001 User's Guide
Enter
Enter
Enter