How To Save And Recall Limits - Keysight Technologies FieldFox A Series User Manual

Table of Contents

Advertisement

——On Pass
——On Fail
——Off
No audible beep sounds.
Warning On Off
points that fail are highlighted.
Limit Type
Then Fixed or Relative.
Offset
Shifts the limit line up or down. Enter an offset value using the rotary knob, the
arrows, or by entering a value with the number keys.
——Enter a positive offset to shift the limit line UP, usually used with UPPER limits.
——Enter a negative offset to shift the limit line DOWN, usually used with LOWER
limits.
Margin
Shifts the limit line up or down like an offset, but only virtually. The limit line does
NOT appear to move. Enter a Margin value using the rotary knob, the
entering a value with the number keys.
Enter a positive margin to virtually shift the limit line UP.
Enter a negative margin to virtually shift the limit line DOWN.

How to Save and Recall Limits

— Press Save/Recall then
used for the Save or Recall. Learn more in
— Then press Limit.
— Then choose from the following:
Save Limits
After limit lines have been defined, this saves the line definition to a file on the
specified device.
Recall Limits
616
Data Analysis Features
All about Limit Lines
Audible beep sounds when a sweep passes.
Audible beep sounds when a sweep fails.
Shows Pass or Fail indication on the display with each sweep. Data
Device
to set the Device (Internal, SD card, or USB flash drive) to be
Recalls a limit Line definition from the specified device.
"Set File Type and Select Device" on page
Keysight N9927-90001 User's Guide
arrows, or by
677.

Advertisement