Pulse Measurements Mode – Option 330
Pulse Measurement Settings
Pulse Measurement Settings
Select Pulse Measurement Mode before making any setting in this section.
How to select Pulse Measurement Mode
— Press Mode
— Then
— Then
Measurement Selection
The pulse measurement results are presented as either a traditional power
meter display or as a trace graph.
How to select the Measurement
— Press Measure 1
— Then choose from the following:
Frequency / Time
How to set Frequency
— Press Freq/Dist
— Then choose from the following:
472
More
if your FieldFox contains more than five measurement modes.
Pulse Measurements
—
Average
Average power (Meter display)
—
Peak
Peak power (Meter display)
—
Pk to Avg
Displays the difference between the above two
measurements (Meter display)
—
Trace Graph
(default) The primary trace (Tr 1) plots pulse data in
amplitude versus time. Optionally, you can enable a secondary
'zoom' trace. Learn how in
— When a 'Meter display' measurement is selected, the following Pulse
Measurement Mode settings are identical to the standard USB Power
Meter Mode settings.
— Learn how to set Scale in
— Learn how to set Relative and Offset Power in
Power Measurements" on page
— Learn how to set Display Units in
— Learn how to set Limits in
"Zoom Window" on page
"Scale" on page
474.
"Relative and Offset
456.
"Display Units" on page
"Limits" on page
458.
Keysight N9927-90001 User's Guide
473.
456.