Changing Frequency Converter Head Settings

If you want to change the frequency converter head settings while using the
VSA, first it is necessary to disconnect from the 89600A VSA:
1. Disconnect from the 89600A VSA software, using the VSA's disconnect
2. Change modes to another frequency converter compatible mode (e.g.,
3. Change the settings on your frequency converter head.
4. Reconnect the FieldFox to the VSA using the VSA's software (
For more information, refer to
and N991xA/3x/A5x/A6xA and SA Option 233)" on page
Keysight N9927-90001 User's Guide
icon (
).
SA/RTSA/IQA/...)
The frequency range in the VSA should always reflect the frequency
converter head settings when you last connected.
"89600 VSA Software Overview (Requires CPU2
).
418.